1010
CLASS II TYPE A2 ANIMAL HANDLING BIOSAFETY CABINET
NU-602
Model Width Size
(nominal) Electrical* (all sizes) Interior Dimensions
(W x D x H) [mm]
Exterior Dimensions
(W x D x H) [mm] Net Weight
NU-602-400 4 ft
[1.2 m]
115V / 60 Hz
E: 230V / 50-60Hz
46
1/2
x 24
1/2
x 26
1/2
[1232 x 622 x 673]
50
1/2
x 31
3/4
x 77
[1283 x 806 x 1956]
780 lbs
[370 kg]
NU-602-500 5 ft
[1.5 m]
58
1/2
x 24
1/2
x 26
1/2
[1486 x 622 x 673]
62
1/2
x 31
3/4
x 77
[1956 x 806 x 1956]
880 lbs
[415 kg]
NU-602-600 6 ft
[1.8 m]
70
1/2
x 24
1/2
x 26
1/2
[1791 x 622 x 673]
74
1/2
x 31
3/4
x 77
[1588 x 806 x 1956]
980 lbs
[460 kg]
Specifications
Standard Features
HEPEX Zero Leak Airflow System
Locking Swivel Casters
(Low Profile,
Stainless Steel)
Electric Auto-lift
Stainless Steel Interior
Removable Work Tray
Liquid Drain Valve
Fluorescent Lighting
12” Access Opening (305 mm)
105 LFPM Air
Barrier (0.53 m/s)
Tempered Glass View Screen
Pre-Filter System
Magnehelic Gauge
Essentials Control System
Optional Features & Accessories
Ultraviolet Light
Decorative Front Panel
Ground Fault Interrupter
Service Valves
Three Optional Configurations
Custom Solutions
*Specify model with appropriate letter suffix for electrical specifications. “NU-602-400E” for 230 VAC / 50 - 60 Hz
Sterile Cage Delivery System
Configuration Options:
Dirty Cage Collection System Waste Collection System
The LabGard® ES (Energy Saver) NU-602 Class II, Type A2 Animal
Handling Biosafety Cabinet is intended to be used for safe animal and
cage handling in the laboratory. The cabinet is designed for maximum
control over product quality and to offer protection to people working
in the lab. The NU-602 was also created with ergonomics and ease-
of-use in mind, so it is easy and convenient for lab technicians to use.
Performance Features:
• 30% exhausted / 70% recirculated
• Dynamic air barrier of 105 fpm (0.53 m/s)
inflow separating laboratory from work
zone environment
• True Laminar Unidirectional Down-flow
air of 75 fpm (0.30 m/s) for minimal
cross contamination
• 250% Filter Load Capacity (±10 Year
Average depending on laboratory
environment)