CLASS II TYPE B2
(100% Exhausted)
ype B2
ecirculated vs. 100%Exhausted
e Air
The LabGard® ES (Energy Saver) TE (Total Exhaust)
NU-560 Class II, Type B2 Biosafety Cabinet is a bench
top model, available with several base stand options for
operation as a full console model. Since the entire work
zone must be 100% exhausted through your facility’s
internal exhaust network, the NU-560 may be used
for work with minute quantities of volatile chemicals
and trace amounts of radionuclides. When used with
proper technique, the LabGard® NU-560 is an effective
primary engineering control. The Type B2 also reduces
the risk of product and personnel exposure to airborne
biological pathogens or particulate chemical agents in
low to moderate risk-hazard environment by creating a
dynamic air barrier at the front of the machine.
The NU-560 also features an exhaust interlock system
that prevents operation of the BSC unless your facility’s
HVAC exhaust flow is sufficient enough to provide the
necessary inflow velocity to maintain the dynamic air
barrier during startup. An externally mounted, energy
efficient LED light fixture provides 860 to 1600 LUX on
your full work surface and comes standard with every
Class II, Type B2 Biosafety Cabinet.
HEPA Filtered Air
Contaminated Work Zone Air
Contaminated Room Air
TouchLink™ Electronic Control System
NU-560 CLASS II TYPE B2 BIOSAFETY CABINET
*Optional IV Bar Shown
25