WFM700/WFM7M Input Module Test Record
WFM700 Series Waveform Monitors Service Manual 2-59
Test Record (cont.)
Instrument Serial Number: Certificate Number:
Input Module Serial Number:
Temperature: Relative Humidity %:
Date of Calibration: Technician:
Performance Test MaximumOutgoingIncomingMinimum
PIX Video Output Level
High Definition
PIX G/Y calculated amplitude error -- 5 % +5%
PIX G/Y black level -- 2 5 m V +25 mV
PIX B/Pb calculated amplitude error -- 5 % +5%
PIX B/Pb black level -- 2 5 m V +25 mV
PIX R/Pr calculated amplitude error -- 5 % +5%
PIX R/Pr black level -- 2 5 m V +25 mV
Standard Definition
PIX G/Y calculated amplitude error -- 5 % +5%
PIX G/Y black level -- 2 5 m V +25 mV
PIX B/Pb calculated amplitude error -- 5 % +5%
PIX B/Pb black level -- 2 5 m V +25 mV
PIX R/Pr calculated amplitude error -- 5 % +5%
PIX R/Pr black level -- 2 5 m V +25 mV
External Reference
1080i 50 100% color bars
Display remains locked to the external reference Pass N/A
1080i 60 100% color bars
Display remains locked to the external reference Pass N/A
720p 59.94 100% color bars
Display remains locked to the external reference Pass N/A
Cable Loss Accommodation
HIgh Definition
No embedded CRC Y or C errors messages Pass N/A
Standard Definition
No embedded CRC Y or C errors messages Pass N/A