
Performance Verification
2-40 WFM700 Series Waveform Monitors Service Manual
10. Set Hi Pass Filter to 1kHz.
11. Change RF generator settings as follows:
LFGEN2 Frequency 1 kHz
12. Repeat steps 7 through 9.
13. Set Hi Pass Filter to 10 kHz.
14. Change RF generator settings as follows:
LFGEN2 Frequency 10 kHz
15. Repeat steps 7 through 9.
16. Change RF generator settings as follows:
LFGEN2 Frequency 100 kHz
17. Set Hi Pass Filter to 1kHz.
18. Adjust the RF generator deviation for a Jitter reading of 4.0 UI.
19. Set Hi Pass Filter to 100 kHz.
20. Repeat steps 7 through 9.