Adjustments
3-14 WFM700 Series Waveform Monitors Service Manual
Adjust signal to cursors,
disregard peaks and dips
Figure 3- 10: Eye vertical scale adjustment screen
16. Touch the Save and Exit soft key to save the new calibration constant.
17. Touch the CONFIG button twice to exit the configuration menu.