SCALE OPERATION
CALIBRATION
Bar Code Calibration Procedure without Remote Display
7. Make sure there is no load on the scale platter. The white LEDs will be
used to indicate the current step in the calibration process.
8. Calibrate a zero load. The white LEDs will blink once periodically. Wait 8
to 10 seconds for scale stability then press the beeper volume switch one
time. The beeper will beep 1 time indicating that the Cal 1 value has been
stored.
Figure 73.
9. Place a half load on the center of the scale platter. The white LEDs will blink
twice periodically. Wait for scale stability, and then press the beeper volume
switch once. The beeper will beep 2 times indicating that the Cal 2 value has
been stored.
Figure 74.
10. Add an additional half load to the existing half to simulate a full load (center
the entire load). The white LEDs will blink three times periodically. Wait for
scale stability, and then press the beeper volume switch once. The beeper
will beep 3 times indicating that the Cal 3 value has been stored.
Figure 75.