SCALE OPERATION
CALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display
7. Make sure there is no load on the scale platter.
8. Calibrate a zero load. Press the right arrow () button once when the
remote display reads CAL 1 and there is no load on the scale platter.
Figure 62.
9. Place a half load (15.00 lbs. or 7.500 kg) on the center of the scale platter
then press the right arrow () button.
Figure 63.
10. Add an additional half load (15.00 lbs. or 7.500 kg) to the existing half to
simulate a full load, center entire load then press the right arrow () button.
Figure 64.
11. Remove half of the load (15.00 lbs. or 7.500 kg), center the remaining load
then press the right arrow () button.